FDA Recall Terminated

Achillon Minimally Invasive Device for Achilles Tendon Rupture Treatment The Achillon Suture System is a single use, sterile device used to treat acute Achilles tendon ruptures. The Achillon Suture System is a mini-invasive procedure indicated to teat acute Achilles tendon ruptures: Acute rupture of the Achill tendon(<10 days); Rupture located between 2 and 8 cm above calcaneum; Open or closed ruptures

Recall: Z-2743-2016 · Initiated June 17, 2016

Recall

Recall Number
Z-2743-2016
Event Number
74613
Firm
Integra LifeSciences Corp.
FEI Number
3003418325
Product Code
NOV
Status
Terminated
Root Cause
Packaging
Initiated
June 17, 2016
Terminated
November 7, 2017
Address
311 Enterprise Dr, Plainsboro, NJ, 08536-3344

Description

Achillon Minimally Invasive Device for Achilles Tendon Rupture Treatment The Achillon Suture System is a single use, sterile device used to treat acute Achilles tendon ruptures. The Achillon Suture System is a mini-invasive procedure indicated to teat acute Achilles tendon ruptures: Acute rupture of the Achill tendon(<10 days); Rupture located between 2 and 8 cm above calcaneum; Open or closed ruptures

Reason

Data from internal studies suggests there may be a potential that the secondary package seal integrity does not provide the level of sterility intended for Minimally Invasive Achilles Tendon Suture System Achillon.

Action

Integra sent an Urgent Voluntary Medical Device Recall letter and recall acknowledgment and return form dated June 17, 2016 to their customers via traceable courier. Customers were asked to review their inventory and complete the attached form. Customers with questions should contact Customer Service at 1-800-654-2873. For questions regarding this recall call 717-840-3431.

Distribution

Nationwide distribution

Quantity

6,729 units