FDA Recall Open, Classified

Orthopedics 6.0MM ROD, COCR, SINGLE HEX, 500MM LENGTH-The RESPONSE" 5.5/6.0 Spine System is intended for immobilization and stabilization of the posterior, non-cervical spine in skeletally mature patients as an adjunct to fusion for the following indications: degenerative disc disease Model Number: 00-1300-6053

Recall: Z-2742-2020 · Initiated July 6, 2020

Recall

Recall Number
Z-2742-2020
Event Number
86027
Firm
OrthoPediatrics Corp
FEI Number
3006460162
Product Code
NKB
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
July 6, 2020
Address
2850 Frontier Dr, Warsaw, IN, 46582-7001

Description

Orthopedics 6.0MM ROD, COCR, SINGLE HEX, 500MM LENGTH-The RESPONSE" 5.5/6.0 Spine System is intended for immobilization and stabilization of the posterior, non-cervical spine in skeletally mature patients as an adjunct to fusion for the following indications: degenerative disc disease Model Number: 00-1300-6053

Reason

Product may have incorrect laser etched alignment lines, the non-conforming laser etch may have compromised strength, lead to unintended fracture of the device

Action

Orthopedics issued Field Action Notice distributed via email on July 6, 2002 stating reason for recall , heath risk and action to take: Distributors are required, to perform sub-recall. The affected product must be returned to OrthoPediatrics Corporation. Replacement product will be provided. Please contact customer service to arrange return of the affected product. Distributors should retrieve affected product that may be with customers. No additional corrective actions need to be taken for product already implanted. Please complete and return the attached response form confirming your understanding of these instructions. Any questions regarding this recall can be directed to Kriss Anderson, VP of Quality Systems and Regulatory Affairs, at OrthoPediatrics at [email protected] or 574-267-0897

Distribution

US Nationwide distribution including in the states of Al, CA, FL, GA, KY, MI, MO, NC, NY, OH, TX.

Quantity

73 units