FDA Recall Terminated

MAKO RIO Standard System, internal cPCI Card CageP/N: 201337. Intended to assist the surgeon in providing software defined spatial boundaries during orthopedic procedures.

Recall: Z-2741-2017 · Initiated June 7, 2017

Recall

Recall Number
Z-2741-2017
Event Number
77553
Firm
Mako Surgical Corporation
FEI Number
3005985723
Product Code
OLO
Status
Terminated
Root Cause
Employee error
Initiated
June 7, 2017
Terminated
June 16, 2020
Address
2555 Davie Rd, Ste 110, Davie, FL, 33317-7424

Description

MAKO RIO Standard System, internal cPCI Card CageP/N: 201337. Intended to assist the surgeon in providing software defined spatial boundaries during orthopedic procedures.

Reason

An intermittent electrical problem that could lead to a loss of system power due to a loose screw connection.

Action

Customers were sent notices titled "Urgent Medical Device Correction Notification" on 06/12/2017 via UPS Next Day Mail. The firm notifies consignees of the reason for recall and instructs that users of the recalled product be provided the recall notice. Users may continue to use their MAKO system; and, Stryker/MAKO Service personnel will contact users o schedule time to inspect and replace the card cage, if necessary. Also, users are instructed to return the completed Business Reply Form within 5 business days. The completed form can be sent to [email protected] or faxed to 855-544-4807. Any questions can be directed to Clayton Odor at 954-628-0502.

Distribution

Distributed to the states of AZ, CA, CO, CT, FL, GA, ID, LA, MA, MO, MS, NC, OH, PA, TN, TX, & WA.

Quantity

20