FDA Recall Terminated

VitalGo Total Lift Bed VG-TLB425T-TQR, Version 71064V0-27003

Recall: Z-2739-2018 · Initiated April 12, 2018

Recall

Recall Number
Z-2739-2018
Event Number
80495
Firm
Vitalgo Inc.
FEI Number
3010034871
Product Code
FNL
Status
Terminated
Root Cause
Component change control
Initiated
April 12, 2018
Terminated
March 12, 2019
Address
3520 NW 56th St, Ste 303, Fort Lauderdale, FL, 33309-2229

Description

VitalGo Total Lift Bed VG-TLB425T-TQR, Version 71064V0-27003

Reason

Potential for unintentional movement of the support surface of the bed from tilted position to horizontal position or any position in between.

Action

The customer was notified of the recall on about 04/12/2018 via letter. The letter was follow up to previous communications on 02/19/2018, 03/06/2018, 03/07/2018, 03/26/2018, and 04/09/2018 and reminded the customer that all affected beds at each service center are to be placed on Quality Hold. Further directions will be provided at a later time to conduct a correction on the beds.

Distribution

US distribution to Texas.

Quantity

85 units