FDA Recall Open, Classified

DR 800. Digital Radiography X-ray System.

Recall: Z-2738-2024 · Initiated July 29, 2024

Recall

Recall Number
Z-2738-2024
Event Number
95115
Firm
Agfa N.V. Septestraat 27 Mortsel Belgium
FEI Number
3001556265
Product Code
JAA
Status
Open, Classified
Root Cause
Device Design
Initiated
July 29, 2024
Posted
August 27, 2024

Description

DR 800. Digital Radiography X-ray System.

Reason

Potential for the front lever chain of the DR 800 table to fail.

Action

Consignees were sent an URGENT FIELD SAFETY NOTICE by email, dated 7/29/24. In the notice, consignees are asked to check the correct tightening of the floor plate fixing bolts if a high and persistent noise is heard from the device during the tilting or elevation movement of the tabletop. If the noise persists, the device must be stopped immediately in a horizontal position at the maximum height. Agfa will coordinate with consignees to schedule an inspection of affected devices and to install hardware intended to reinforce equipment before 10/15/24. The recall notification is to be shared within consignee organizations and forwarded to those devices were transferred to. Consignees are asked to maintain awareness of this notice until devices are corrected. Completed response forms are to be provided via email to [email protected]. Consignees with any questions are to call 1-877-777-2432 and reference PRB2001106.

Distribution

Worldwide - US Nationwide distribution in the states of CA, FL, IL, IN, KS, LA, MA, MO, NJ, NC, OH, OK, SC, VA, WV, WI; and the countries of AU, BE, BW, CA, CO, CZ, DE, ES, FR, GB, GR, IL, IN, IT, KZ, MX, PL, RS, UZ, ZA.

Quantity

141 units