2.4, 3.0, and 4.0 mm Gridlock Screw Driver Bit, Model Number 310-30-003, Lot Number TSL002605, NON-STERILE. For use with the Gridlock Plating System for trauma and reconstructive procedures of small bones.
Recall
- Recall Number
- Z-2738-2015
- Event Number
- 71936
- Firm
- Trilliant Surgical Ltd.
- FEI Number
- 3007420745
- Product Code
- HRS
- Status
- Terminated
- Root Cause
- Component design/selection
- Initiated
- July 31, 2015
- Posted
- September 4, 2015
- Terminated
- January 26, 2017
- Address
- 6721 Portwest Dr, Ste 160, Houston, TX, 77024-8019
Description
2.4, 3.0, and 4.0 mm Gridlock Screw Driver Bit, Model Number 310-30-003, Lot Number TSL002605, NON-STERILE. For use with the Gridlock Plating System for trauma and reconstructive procedures of small bones.
Gridlock screw driver bits for the Gridlock Plating System were identified to be out of specification and have the potential for shallow engagement between the driver and the screw, resulting in an increased likelihood of stripping screw heads intraoperatively. Specifically, the minor diameter of the drive feature is above the specification, potentially reducing the engagement between the driver and the screw head.
The firm sent out a written notification of the recall on 7/31/2015.
NY, OH, CA, SC, NJ, AZ, VA, IN, FL, MN, OK, AR, MO, WI, GA, SD, IA, IL, TX
108 units