FDA Recall Terminated

SPECTRUM Pump, Model No. 35700BAX. Intended to be used for the controlled administration of intravenous fluids.

Recall: Z-2738-2014 · Initiated September 3, 2014

Recall

Recall Number
Z-2738-2014
Event Number
69122
Firm
Baxter Healthcare Corp.
FEI Number
1416980
Product Code
FRN
Status
Terminated
Root Cause
Equipment maintenance
Initiated
September 3, 2014
Posted
October 1, 2014
Terminated
March 16, 2017
Address
25212 W. Illinois Route 120, Round Lake, IL, 60073-9799

Description

SPECTRUM Pump, Model No. 35700BAX. Intended to be used for the controlled administration of intravenous fluids.

Reason

One Service Technician may not have correctly serviced specific Sigma Spectrum Infusion Pumps according to established procedures during the time period of 5/5/2014 through 6/3/2014.

Action

Initial notification was initiated via telephone call by the Baxter Medina Service Center on 8/28/14 to all affected Sigma SPECTRUM Infusion Pump consignees. URGENT DEVICE CORRECTION Letters (dated 9/03/2014) were sent to the consignees via USPS first class mail on 9/03/14 formally informing them of the recall. The letters identified the affected product, the description of the issue, the hazard involved, as well as, the actions to be taken by customer. Customers are being instructed to immediately remove the pump from use and return the pump to Baxter for inspection. The firm will provide replacements. Customers are to contact Baxter Healthcare Medina at 800-356-3454 for technical questions regarding the letter.

Distribution

Worldwide Distribution -- USA and Canada.

Quantity

USA: 56 units, Canada: 2 units