SPECTRUM Pump, Model No. 35700BAX. Intended to be used for the controlled administration of intravenous fluids.
Recall
- Recall Number
- Z-2738-2014
- Event Number
- 69122
- Firm
- Baxter Healthcare Corp.
- FEI Number
- 1416980
- Product Code
- FRN
- Status
- Terminated
- Root Cause
- Equipment maintenance
- Initiated
- September 3, 2014
- Posted
- October 1, 2014
- Terminated
- March 16, 2017
- Address
- 25212 W. Illinois Route 120, Round Lake, IL, 60073-9799
Description
SPECTRUM Pump, Model No. 35700BAX. Intended to be used for the controlled administration of intravenous fluids.
One Service Technician may not have correctly serviced specific Sigma Spectrum Infusion Pumps according to established procedures during the time period of 5/5/2014 through 6/3/2014.
Initial notification was initiated via telephone call by the Baxter Medina Service Center on 8/28/14 to all affected Sigma SPECTRUM Infusion Pump consignees. URGENT DEVICE CORRECTION Letters (dated 9/03/2014) were sent to the consignees via USPS first class mail on 9/03/14 formally informing them of the recall. The letters identified the affected product, the description of the issue, the hazard involved, as well as, the actions to be taken by customer. Customers are being instructed to immediately remove the pump from use and return the pump to Baxter for inspection. The firm will provide replacements. Customers are to contact Baxter Healthcare Medina at 800-356-3454 for technical questions regarding the letter.
Worldwide Distribution -- USA and Canada.
USA: 56 units, Canada: 2 units