FDA Recall Terminated

Mindray DPM 7 Monitor, Multi Parameter Patient Monitor (with Arrhythmia Detection and Alarms) Intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters including ECG (3-lead or 5-lead selectable), arrhythmia detection, ST Segment analysis, and heart rate.

Recall: Z-2737-2015 · Initiated July 27, 2015

Recall

Recall Number
Z-2737-2015
Event Number
71934
Firm
Mindray DS USA, Inc. dba Mindray North America
FEI Number
2221819
Product Code
MHX
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
July 27, 2015
Posted
September 4, 2015
Terminated
May 2, 2017
Address
800 Macarthur Blvd, Mahwah, NJ, 07430-2001

Description

Mindray DPM 7 Monitor, Multi Parameter Patient Monitor (with Arrhythmia Detection and Alarms) Intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters including ECG (3-lead or 5-lead selectable), arrhythmia detection, ST Segment analysis, and heart rate.

Reason

The DPM 7 Monitors may display a black screen.

Action

Mindray sent an Urgent Medical Device Correction / DPM 7 Monitor letter, dated July 27, 2015, to their affected customers via certified mail and return receipt. The letter identified the problem as well as the affected product, adverse effects on patients, and corrective action. The firm will repair the potentially affected monitor at no cost to the customer. Customers are to contact their Mindray Service Representative at 1-800-288-2121 to arrange for the repair.

Distribution

Distributed to the states of CT, IA, KY, MS, PA, UT and WA.

Quantity

13 units