FDA Recall Terminated

System 1 Base 110V, Catalog 801763, Terumo Cardiovascular Systems, Corp., Ann Arbor, MI The product is indicated for up to 6 hours of extracorporeal circulation of blood for arterial perfusion, regional perfusion and cardiopulmonary bypass procedures when used by a qualified medical professional

Recall: Z-2735-2011 · Initiated March 31, 2011

Recall

Recall Number
Z-2735-2011
Event Number
58547
Firm
Terumo Cardiovascular Systems Corporation
FEI Number
1828100
Product Code
DTQ
Status
Terminated
Root Cause
Device Design
Initiated
March 31, 2011
Posted
July 6, 2011
Terminated
August 28, 2013
Address
6200 Jackson Road, Ann Arbor, MI, 48103-9586

Description

System 1 Base 110V, Catalog 801763, Terumo Cardiovascular Systems, Corp., Ann Arbor, MI The product is indicated for up to 6 hours of extracorporeal circulation of blood for arterial perfusion, regional perfusion and cardiopulmonary bypass procedures when used by a qualified medical professional

Reason

The firm has received multiple reports of the lower roller pump tube clamp mechanism failing to function in a small population of devices, making it difficult to remove or insert the tubing. If the failure occurs, the clamp dial will no longer be able to control opening/closing of the tube clamp mechanism.

Action

Terumo Cardiovascular Systems sent an URGENT MEDICAL DEVICE CORRECTION letter to all affected customers dated March 21, 2011. The letter identified the product, the problem and the action needed to be taken by the customer. An additional URGENT MEDICAL DEVICE RECALL correction notification and cover letter was sent to all consignees on XX/XX/XX to explain the reason for the expansion of the recall and will provide an update to the timing of implementation of the field correction. Expansion now includes devices manufactured between 2006 and March 2012. The correction strategy is to issue the customers a letter and addendum to the Operator's Manual that specifies what the user should do in the event of a tube clamp failure for any reason. The letter states that the firm is still investigating the issue and the correction. The letter states that in the coming months when a resolution is identified, consignees may be contacted to make arrangements for a upgrade to the system. There is no specific timeline for the correction to be made. If you have any further questions, please call a Terumo CVS Customer Service at 1-800-521-2818. Customer Service hours Monday - Friday, 8 Am - 6 Pm EST.

Distribution

Worldwide Distribution -- USA (nationwide) and the countries of Argentina, Australia, Belgium, Canada, Chile, Columbia, Costa Rica, Dominican Republic, Egypt, Germany, Guatemala, Honduras, Hong Kong, Indonesia, Japan, Jordan, Korea, Kuwait, Malaysia, Mexico, Pakistan, Republic of Georgia, Russia, Saudi Arabia, Singapore, South Africa, Taiwan, Thailand, The Phillipines, Turkey, United Arab Emirates and Vietnam.

Quantity

984 devices