Waterpik Sonic-Fusion Professional, Model: SF-02 Product Usage: The device is composed of a unit base and an oral irrigator/ battery operated brush power handle. The device provides a pulsating stream of potable water and/or powered toothbrush action to remove plaque and food particles from between the teeth and below the gum line. The power handle can also function as a rechargeable battery powered mechanical toothbrush.
Recall
- Recall Number
- Z-2734-2018
- Event Number
- 80474
- Firm
- Water Pik, Inc.
- FEI Number
- 1712259
- Product Code
- JET
- Status
- Terminated
- Root Cause
- Under Investigation by firm
- Initiated
- July 2, 2018
- Terminated
- May 24, 2022
- Address
- 1730 East Prospect Road Fort, Collins, CO, 80553-0001
Description
Waterpik Sonic-Fusion Professional, Model: SF-02 Product Usage: The device is composed of a unit base and an oral irrigator/ battery operated brush power handle. The device provides a pulsating stream of potable water and/or powered toothbrush action to remove plaque and food particles from between the teeth and below the gum line. The power handle can also function as a rechargeable battery powered mechanical toothbrush.
Charging base may overheat with localized melting and sparking, possibly causing fire, shock or burns.
On 07/02/2018, the firm sent customers an Urgent Medical Device Recall notice. Customers were instructed to stop using affected devices, immediately unplug the units, and to return these devices using firm-provided prepaid return labels. For questions, consumers should call 1-800-674-7718 from 7:00 a.m. to 5:00 p.m. MDT, Monday through Friday. On 07/05/2018, Urgent Medical Device Recall notices were mailed to all customers via overnight mail. On 07/13/2018, the firm issued a press release, which also stated to call 1-800-674-7718 to receive a product return kit. Customers were encouraged to report adverse events to MedWatch via http://www.fda.gov/MedWatch/report.htm (online), http://www.fda.gov/MedWatch/getforms.htm (mail), 1-800-332-1088 (phone), or 1-800-FDA-0178 (fax).
U.S. Nationwide Distribution in the states of TX, KY, WA, NY, VA, CA, NC, IA, MA, OR, OH, GA, NV, CT, IL, SC, PA, MD, IN, WV, ND, MS, CO, AZ, WI, KS, FL, HI, OK, VT, MO, ID, RI, MN, MI, AR, LA, TN, DC, NJ, NH, NE, AL, UT, ME, NM, AK, DE, SD, MT, WY and country of Canada