FDA Recall Terminated

Access Thyroglobulin Assay; 100 determinations, 2 packs, 50 tests/pack. For the quantitative determination of thyroglobulin levels in human serum and plasma using the Access Immunoassay Systems.

Recall: Z-2733-2015 · Initiated July 28, 2015

Recall

Recall Number
Z-2733-2015
Event Number
71878
Firm
Beckman Coulter Inc.
FEI Number
2050012
Product Code
MSW
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
July 28, 2015
Posted
September 3, 2015
Terminated
April 13, 2016
Address
250 S Kraemer Blvd, Brea, CA, 92821-6232

Description

Access Thyroglobulin Assay; 100 determinations, 2 packs, 50 tests/pack. For the quantitative determination of thyroglobulin levels in human serum and plasma using the Access Immunoassay Systems.

Reason

Beckman Coulter, Inc. is recalling Access Thyroglobulin reagent packs because some may exhibit greater than 10% imprecision at concentrations <10 ng/mL.

Action

On 07/28/2015 the firm sent out customer notification letters via first class mail and e-mail. The firm's letter contained the following instructions: 1. Discard all Access Thyroglobulin reagent pack lots listed in the letter. 2. Evaluate Access Thyroglobulin results <10 ng/mL when using the reagent pack lot numbers listed to determine if controls and/or patients were higher than expected or did not match the clinical status of the patient. 3. At the discretion of the Laboratory Director, notify clinicians that it is possible the Access Thyroglobulin sample results reported by the laboratory were affected by this issue. The firm's notification letter requests that the information be shared with laboratory staff. If any affected products have been forwarded provide a copy of the letter to them. The firm requests that customers complete and return the Response Form within 10 days. Any questions contact the Customer Care Center: 1. From website at http://www.beckmancoulter.com 2. By phone: call 1-800-854-3633 in the United States and Canada. 3. Outside the United States and Canada, contact local Beckman Coulter representative

Distribution

Worldwide Distribution-US (nationwide) including Puerto Rico and the countries of Algeria, Australia, Austria, Belgium, Bolivia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Czech Republic, Egypt, France, Germany, Greece, Honduras, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Korea, Republic of, Lebanon, Lithuania, Mexico, Morocco, New Zealand, Oman, Philippines, Portugal, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, Vietnam and Yemen.

Quantity

3,035 (U.S.) and 7,123 (International)