FDA Recall Terminated

UniCel DxH 800 and DxH 600 Coulter Cellular Analysis System. Catalog No. 629029, B24465, B24802, B68304 (Unicel DxH 800 Coulter Cellular Analysis System). Catalog No. B23858 (Unicel DxH600 Coulter Cellular Analysis System).

Recall: Z-2732-2016 · Initiated August 8, 2016

Recall

Recall Number
Z-2732-2016
Event Number
74912
Firm
Beckman Coulter Inc.
FEI Number
2050012
Product Code
GKZ
Status
Terminated
Root Cause
Software design
Initiated
August 8, 2016
Posted
September 2, 2016
Terminated
April 29, 2017
Address
250 S Kraemer Blvd, Brea, CA, 92821-6232

Description

UniCel DxH 800 and DxH 600 Coulter Cellular Analysis System. Catalog No. 629029, B24465, B24802, B68304 (Unicel DxH 800 Coulter Cellular Analysis System). Catalog No. B23858 (Unicel DxH600 Coulter Cellular Analysis System).

Reason

There is a possibility of specimen misidentification for the stand-alone DxH 800 and DxH 600 where the Primary Identifier is configured to Tube Position ID.

Action

An Urgent Medical Device Recall letter dated 8/8/16 was sent to all customers who purchased the Unicel DxH 800 and DxH 600 Coulter Cellular Analysis System to inform them that there is a possibility of specimen misidentification for the stand-alone DxH 800 and DxH 600 where the Primary Identifier is configured to Tube Position ID. Customers are instructed to complete and return the enclosed response form within 10 days. Customers with any questions regarding the notice are instructed to contact Customer Support Center: http://www.beckmancoulter.com/customersupport/support; 800-526-7694 in the US and Canada; outside the US and Canada, contact local Beckman Coulter representative.

Distribution

Affected geographies: Andorra, Australia, Belarus, Belgium, Bosnia and Herzegovina, Brunei Darussalam, Canada, China, Columbia, Croatia, Czech Republic, Egypt, Ecuador, France, Germany, Gibraltar, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Kuwait, Lebanon, Macao, Malaysia, Mayotte, Mexico, Morocco, Myanmar, New Zealand, Norway, Oman, Pakistan, Panama, Paraguay, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Republic, United Kingdom, Uruguay and Vietnam.

Quantity

5,282 units total (2,877 units in US)