FDA Recall Terminated

DYONICS POWER Shaver Blades, 4.5mm Long Full Radius, package of 3 Catalog Number: 7205341 Product Usage: Smith & Nephew DYONICS Disposable Endoscopic Blades are indicated for resection of soft and osseous tissues in both large and small articular cavities.

Recall: Z-2732-2011 · Initiated May 6, 2011

Recall

Recall Number
Z-2732-2011
Event Number
58733
Firm
Smith & Nephew, Inc. Endoscopy Division
FEI Number
3003604053
Product Code
HAB
Status
Terminated
Root Cause
Process control
Initiated
May 6, 2011
Posted
July 1, 2011
Terminated
May 9, 2012
Address
150 Minuteman Drive, Andover, MA, 01810-1031

Description

DYONICS POWER Shaver Blades, 4.5mm Long Full Radius, package of 3 Catalog Number: 7205341 Product Usage: Smith & Nephew DYONICS Disposable Endoscopic Blades are indicated for resection of soft and osseous tissues in both large and small articular cavities.

Reason

Product has insufficient weld between the shaft tube and tip, could result in the tip separating from the shaft tube during use.

Action

Smith & Nephew notified customers by an "URGENT - PRODUCT RECALL 1st NOTIFICATION" letter dated May 6, 2011. The letter describes the product, problem, and actions to be taken. The letter asked customers to call 800-343-5717 (option 3) or send an email to [email protected] for a Return Authorization (RA) Number to return product, obtain credit or order replacement blades. Customers are to complete the form and return to the address provided being sure to reference the RA number. For any questions regarding this recall customers were instructed to call 508-337-3731.

Distribution

Worldwide Distribution -- USA (nationwide) and countries of: Argentina, Australia, Brazil, Canada, Chile, China, Columbia, Germany, Hong Kong, Italy, Japan, Mexico, Netherlands, Turkey, and United Kingdom.

Quantity

555 units