FDA Recall Terminated

Beckman Coulter Synchron AQUA CAL Level 1, Part No. 471288 Product Usage: Used for the system calibration of Na, K, Cl, Urea N, Urea, Glucose Creatine, Ca, CO2, and P in UniCel DxC Systems.

Recall: Z-2730-2017 · Initiated May 15, 2017

Recall

Recall Number
Z-2730-2017
Event Number
77364
Firm
Beckman Coulter Inc
FEI Number
2050012
Product Code
JIX
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
May 15, 2017
Posted
July 13, 2017
Terminated
September 9, 2020
Address
250 S Kraemer Blvd, Brea, CA, 92821-6232

Description

Beckman Coulter Synchron AQUA CAL Level 1, Part No. 471288 Product Usage: Used for the system calibration of Na, K, Cl, Urea N, Urea, Glucose Creatine, Ca, CO2, and P in UniCel DxC Systems.

Reason

The firm identified a negative shift of 5mmol/L for sodium recovery in Synchron Control Level 3 when calibrated with AQUA Cal 1.

Action

Recall notices will be sent via mail or email to customers the week of 05/15/2017. Customers are instructed to discard the affected calibrator and request a replacement. Call Client Services at (800) 526-3821 (US only)

Distribution

US Nationwide Distribution in the states of AL, AR, AK, CA, CO, DE, FL, GA, ID, IL, IN, IO, KS, KY, LA, MA, MD, MS, MI, ME, MO, NE, NV, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WV, WI, and Puerto Rico Brunei, Canada, China, Czech Republic, France, Guatemala, Hong Kong, India, Ireland, Italy, Japan, Lebanon, Mexico, Netherlands, Panama, Paraguay, Philippines, Qatar, South Africa, Taiwan, Thailand, Turkey, United Arab Emirates, & United Kingdom.

Quantity

4444 AQUA CAL Level 1 kits