FDA Recall Terminated

Plum A+3 Infusion Pump. List Number 12618. The Plum A+ Infusion Systems are a multi-purpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care. The Plum A+ family infusion pumps are self-contained, microprocessor based infusion devices.

Recall: Z-2729-2016 · Initiated February 6, 2013

Recall

Recall Number
Z-2729-2016
Event Number
64261
Firm
Hospira Inc.
FEI Number
3013319212
Product Code
FRN
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
February 6, 2013
Posted
September 6, 2016
Terminated
November 14, 2016
Address
275 N Field Dr, Lake Forest, IL, 60045-2579

Description

Plum A+3 Infusion Pump. List Number 12618. The Plum A+ Infusion Systems are a multi-purpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care. The Plum A+ family infusion pumps are self-contained, microprocessor based infusion devices.

Reason

The fluid shield diaphragm for the Plum A+ Infusion pumps was undersized, possibly resulting in alarms and delay or interruption of therapy.

Action

Hospira sent an Urgent Device Field Correction letter, dated February 6, 2013, to customers. Customers were requested to return a reply form to Stericycle indicating that they have received the letter, and notified their users and/or customers. The customers were asked to supply the number of Plum A+ infusers at their facility. Product will be corrected by Hospira in the field.

Distribution

United States nationwide distribution including PR, and USVI. Costa Rica, Turks and Caicos, and Dominican Republic.

Quantity

357,778 total