FDA Recall Open, Classified

Ion Endoluminal System, REF: 380748-65

Recall: Z-2720-2024 · Initiated July 12, 2024

Recall

Recall Number
Z-2720-2024
Event Number
95046
Firm
Intuitive Surgical, Inc.
FEI Number
3001675293
Product Code
EOQ
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
July 12, 2024
Posted
August 23, 2024
Address
1266 Kifer Rd, Sunnyvale, CA, 94086-5304

Description

Ion Endoluminal System, REF: 380748-65

Reason

There is a potential that the robotic system's instrument cart arm may contain screws that may potentially fail which could result in uncontrolled motion of a catheter inside a patient's airways.

Action

On July 12, 2024, Intuitive issued a "Urgent Medical Device Correction" notification to affected consignees via E-Mail. 1. This notice needs to be passed on to all those who need to be aware within your organization or functions where the potentially affected devices have been transferred. 2. Please retain a copy of this letter, place a copy with your affected system, and keep the acknowledgement form for your files. 3. Complete the attached Acknowledgement Form immediately and return it via fax or email to Intuitive as instructed on the form. 4. An Intuitive Representative will schedule a site visit to perform replacement of the affected FIM assembly. Once replaced by the Intuitive Representative the Ion System can be used as designed. 5. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, regular mail, or fax.

Distribution

US Nationwide distribution in the states of AZ and CA.

Quantity

3 units