FDA Recall Terminated

ECMO Heater Model 333, Cardiopulmonary bypass temperature controller. This device is intended to provide temperature-controlled warm water flow to a blood heat exchanger in order to maintain a patient's blood temperature. It can be used during extracorporeal circulation procedures. It is intended to use with commercially available heat exchangers.

Recall: Z-2720-2011 · Initiated May 2, 2011

Recall

Recall Number
Z-2720-2011
Event Number
58971
Firm
Cincinnati Sub-Zero Products Inc
FEI Number
1516825
Product Code
DWJ
Status
Terminated
Root Cause
Device Design
Initiated
May 2, 2011
Posted
July 1, 2011
Terminated
March 7, 2013
Address
12011 Mosteller Rd, Cincinnati, OH, 45241

Description

ECMO Heater Model 333, Cardiopulmonary bypass temperature controller. This device is intended to provide temperature-controlled warm water flow to a blood heat exchanger in order to maintain a patient's blood temperature. It can be used during extracorporeal circulation procedures. It is intended to use with commercially available heat exchangers.

Reason

The probe jack contacts (terminals) on certain of the firm's ECMO Heater devices are located too close to the solid state relay contacts (terminals) which can lead to the terminals coming into contact with each other. If this situation occurs, the service technician, user, and/or patient could be exposed to electric shock during servicing and/or use.

Action

Cincinnati Sub-Zero Products Inc issued an "Urgeant Medical Device Field Action" letter on May 2, 2011 to all affected customers. The letter includes; description of product and problem, affected lots and instructs customers to discontinue its use. The product is to be returned to the Service Department for an update and the customer is instructed to complete and mail Field Action Form. For additional contact the firm at 1-800-989-7373 or (513) 772-8810.

Distribution

Worldwide Distribution and countries of The Netherlands, Kuwait, India, Saudi Arabia, Taiwan and Australia. (USA) Nationwide including the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IL, IN, LA, MA, MD, MI, MIN, MO, NC, NE, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, WA, and WI.

Quantity

116 units