FDA Recall Open, Classified

Nanoplasty 3D Bunion Correction System Implant, Model Number: SK57. System contains the 01-00066 1st Metatarsal Plate as a component.

Recall: Z-2718-2024 · Initiated June 2, 2024

Recall

Recall Number
Z-2718-2024
Event Number
94977
Firm
Treace Medical Concepts, Inc.
FEI Number
3011623994
Product Code
HRS
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
June 2, 2024
Posted
August 23, 2024
Address
100 Palmetto Park Pl, Ponte Vedra, FL, 32081-6202

Description

Nanoplasty 3D Bunion Correction System Implant, Model Number: SK57. System contains the 01-00066 1st Metatarsal Plate as a component.

Reason

Potential issue causing a higher frequency of interference with locking screws and/or inserter rods than anticipated, resulting in screws stripping and/or not fully seating in the plates and/or interface issues between inserter rods and plates.

Action

Distributors were contacted via phone or in person to hold product starting on 06/02/2024. Recall notifications were disseminated on 06/13/2024 via email containing the recall letter, acknowledgement, and shipping label for return. On 07/29/2024, Treace Medical contacted surgeons via letter. Surgeons were instructed to monitor patients for adverse events and to complete and return the acknowledgement form.

Distribution

US Nationwide distribution in the states of CO, FL, MD, MI, NC, SC, TX, VA.

Quantity

40 kits containing 40 plates