ArjoHuntleigh Flowtron ACS900, medical pump, software V1.099 Device is used to prevent the occurrence of Deep Vein Thrombosis (DVT) in at-risk patients. The system must be used as part of a prescribed plan of car under the supervision of trained medical and/or Clinical staff. The system is intended for use only in professional healthcare facilities. It is not intended to be used in the home healthcare environment.
Recall
- Recall Number
- Z-2714-2017
- Event Number
- 77551
- Firm
- Arjo, Inc. dba ArjoHuntleigh
- FEI Number
- 1419652
- Product Code
- JOW
- Status
- Terminated
- Root Cause
- Under Investigation by firm
- Initiated
- June 14, 2017
- Posted
- June 26, 2017
- Terminated
- August 12, 2022
- Address
- 2349 W Lake St, Addison, IL, 60101-6183
Description
ArjoHuntleigh Flowtron ACS900, medical pump, software V1.099 Device is used to prevent the occurrence of Deep Vein Thrombosis (DVT) in at-risk patients. The system must be used as part of a prescribed plan of car under the supervision of trained medical and/or Clinical staff. The system is intended for use only in professional healthcare facilities. It is not intended to be used in the home healthcare environment.
Reports of the device failure involving delivering continuous pressure to the calf garment and not deflating as intended, without triggering an alarm.
ArjoHuntleigh sent a Field Safety Notice and Customer Response form to affected customers describing the reason for the recall and the actions to be taken. The instruction to customers is to review the Field Safety Notice and complete the Customer Response Form to acknowledge receipt and understanding of the Field Safety Notice and the required actions. These actions taken are as follows: 1. Ensure that all caregivers and users of the ArjoHuntleigh Flowtron ACS900 pump are made aware of this Field Safety Notice (FSN) and all listed devices at facility are available to be upgraded to software version V2.000 during the service technician visit. 2. Complete and sign the enclosed Customer Response Form and return this form to the local ArjoHuntleigh office. Note: ArjoHuntleigh will contact the designated facility representative listed on the Customer Response Form to schedule service to upgrade the software on all listed devices at your facility, free of charge. The device may stay in use until the upgrade will be performed, on the condition that the patients limbs are monitored frequently and the garments are checked to ensure that they are correctly fitted to the patient and that deflation of the garments is occurring on a regular basis. It is also recommended that the pump LCD display is regularly checked to ensure that it shows that the garments are inflating and deflating correctly (in the limited number of complaints, the LCD display was blank). This Field Safety Notice must be distributed to those individuals who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Additionally they should maintain awareness of the notice and resulting action for an appropriate period to ensure effectiveness of the corrective action. The letter states that if the customer has additional questions or they require assistance completing the Customer Response Form, they should contact ArjoHu
Worldwide Distribution - US (Nationwide) and internationally to the following countries: United Kingdom, Australia, New Zealand, Japan, Italy, United Arab Emirates, France, Canada, Brazil, Austria, Saudi Arabia, Qatar, India, Switzerland, Bahrain, Kuwait, South Africa, Lebanon, Netherlands, Poland, Spain, Ecuador, Hong-Kong, Colombia, Malaysia, Paraguay, Mauritius, Cyprus, Greece, Oman, Norway, Ireland, Pakistan, Czech Republic, Mexico, Sweden, Tanzania, Hungary, Singapore, Jordan, Israel, Denmark, Chile, and Germany.
40,675 units worldwide