FDA Recall Terminated

The 9025TRU Arterial Blood Gas (ABG) Kit is used to obtain arterial blood sampling for pH, Blood Gas, Electrolyte and Metabolite Analysis.

Recall: Z-2713-2016 · Initiated June 27, 2016

Recall

Recall Number
Z-2713-2016
Event Number
74706
Firm
Cardinal Health
FEI Number
3001236905
Product Code
CBT
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
June 27, 2016
Terminated
March 2, 2018
Address
1300 Waukegan Rd, Waukegan, IL, 60085-6724

Description

The 9025TRU Arterial Blood Gas (ABG) Kit is used to obtain arterial blood sampling for pH, Blood Gas, Electrolyte and Metabolite Analysis.

Reason

Arterial Blood Gas (ABG) Kit- Missing Safety Component. The ABG Kit(s) have been identified as missing the locking clasp component. CareFusion/BD has identified a potential customer/user safety risk associated with the missing locking clasp component.

Action

CareFusion/BD sent out a FIELD SAFETY NOTICE of the field correction in writing on June 27, 2016. The letter described the nature of the recall and provided actions to be performed. Carefusion/BD directed customers to take the following actions: Step #1: Inspect current inventory on-hand to identify the presence of the affected Arterial Blood Gas Sampling Kit(s). Step #2: Complete the Customer Response Form by checking all appropriate boxes and providing all accurate information as required. Step #3: Destroy all affected product(s) in-stock in accordance with their facility's destruction protocol. Step #4: Return the Customer Response Form to [email protected] Step #5: If the customers want to obtain credit and/or replacement of destroyed product(s) they are asked to call their CareFusion/BD Customer Support Department at 800.323.9088 (Option #1 ). For any additional questions and support concerning the voluntary Field Safety Corrective Action, customers are asked to call (847)362.9303.

Distribution

Worldwide Distribution - US (nationwide) and Canada.

Quantity

Lot #884993 -16,000 units, Lot #903136 -1,900 units and Lot #903137 - 300 units.