FDA Recall Terminated

SMR allen wrench 5 mm, product code 9013.50.210

Recall: Z-2707-2016 · Initiated July 21, 2016

Recall

Recall Number
Z-2707-2016
Event Number
74827
Firm
Limacorporate S.p.A Via Nazionale 52 San Daniele Del Friuli Italy
FEI Number
3008021110
Product Code
LXH
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
July 21, 2016
Posted
August 31, 2016
Terminated
January 13, 2017

Description

SMR allen wrench 5 mm, product code 9013.50.210

Reason

Product found to be dimensionally non-conforming (slightly over dimensioned on the diameter of the Zimmer connection) and non-functional for connecting with some T-handles.

Action

The firm received two complaints from the US field about the inta-operative malfunction with the SMR allen wrench 5mm (product code 9013.50.210) with the T-handle. Both parts involved were manufactured with the same lot #14AA311. No adverse effects for patients reported.

Distribution

US, Italy, Australia

Quantity

76