FDA Recall Terminated

rHead Standard Extended Stem, 6mm Collar, Size 4. Orthopedic implant for replacement of the proximal end of the radius in the Stryker rHead Radial Head and Uni-Elbow system.

Recall: Z-2706-2018 · Initiated October 24, 2017

Recall

Recall Number
Z-2706-2018
Event Number
79315
Firm
Howmedica Osteonics Corp.
FEI Number
3003070421
Product Code
JDB
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
October 24, 2017
Terminated
May 18, 2023
Address
325 Corporate Dr, Mahwah, NJ, 07430-2006

Description

rHead Standard Extended Stem, 6mm Collar, Size 4. Orthopedic implant for replacement of the proximal end of the radius in the Stryker rHead Radial Head and Uni-Elbow system.

Reason

Revision rate identified as higher than other proximal radial head replacement devices based on literature review. Product may not perform at expected level post-surgery.

Action

On October 27, 2017 an URGENT MEDICAL RECALL NOTIFICATION was issued to customers titled Product Field Action PR 1642784. The letter requests that recipients notify all necessary individuals within their organization of the recall and that a Stryker trauma representative will collect any affected product from the customer location. Questions or concerns can be directed to 201-831-6693

Distribution

Distribution US Nationwide.

Quantity

36