FDA Recall Open, Classified

Computed Tomography X-ray System; Model uCT 760; Rx; UDI: (01) GTIN: 06971576831043 - Product Usage: intended to produce cross-sectional images of the body using computer reconstruction of x-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, cardiac, and vascular).

Recall: Z-2703-2020 · Initiated June 20, 2020

Recall

Recall Number
Z-2703-2020
Event Number
85868
Firm
Shanghai United Imaging Healthcare Co., Ltd. No. 2258 Chengbei Rd, Jiading Ind. Shanghai China
FEI Number
3011015597
Product Code
JAK
Status
Open, Classified
Root Cause
Device Design
Initiated
June 20, 2020
Posted
July 27, 2020

Description

Computed Tomography X-ray System; Model uCT 760; Rx; UDI: (01) GTIN: 06971576831043 - Product Usage: intended to produce cross-sectional images of the body using computer reconstruction of x-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, cardiac, and vascular).

Reason

The metal edge overlap of the mylar strip in the gantry was uneven, which may create a sharp edge.

Action

A Customer Notification Letter was sent to the affected consignees on 06/20/2020 via email. United Imaging will replace the mylar strip with an improved one on the affected products via on-site update. On 8/30/2024, a correction notice was sent identifying an additional serial number.

Distribution

US Nationwide distribution including in the states of CA, IN and TX. See uploaded distribution list for more details.

Quantity

5 units