FDA Recall Terminated

Oxford Femoral Knee System REF 154600, LOT 2272736, WARNING DO NOT RESTERILIZE, STERILE UNLESS PACK DAMAGED, BIOMET UK Ltd, (510K# K011138) This device is used in partial knee replacement surgery as a result of patients who are experiencing degenerative arthritis.

Recall: Z-2698-2011 · Initiated April 22, 2011

Recall

Recall Number
Z-2698-2011
Event Number
58702
FEI Number
3002806535
Product Code
HRY
Status
Terminated
Root Cause
Packaging
Initiated
April 22, 2011
Posted
June 29, 2011
Terminated
August 3, 2011
Address
Biomet U.K., Ltd. Waterton Industrial Estate Bridgend, South Wales United Kingdom

Description

Oxford Femoral Knee System REF 154600, LOT 2272736, WARNING DO NOT RESTERILIZE, STERILE UNLESS PACK DAMAGED, BIOMET UK Ltd, (510K# K011138) This device is used in partial knee replacement surgery as a result of patients who are experiencing degenerative arthritis.

Reason

This recall was initiated due to a customer complaint which reported that while peeling off the Tyvek lid on the blister pack, very little effort was required to remove it and this could lead to contamination of the device.

Action

Biomet sent a "FIELD SAFETY NOTICE" dated April 21, 2011, to all affected accounts. The notice identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue use of the affected product, locate and remove from inventory and quarantine pending return to their Bioment distributor. Customers were also instructed to complete and return the attached "Fax Back Response Form" to their local Biomet distributor. For questions customers should contact their local Biomet Contact.

Distribution

Outside the US to Austria, Belgium, and Germany

Quantity

12 devices