FDA Recall Open, Classified

IBA Proton Therapy System - PROTEUS 235- Designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation

Recall: Z-2693-2024 · Initiated July 10, 2024

Recall

Recall Number
Z-2693-2024
Event Number
95024
Firm
Ion Beam Applications S.A. Chemin Du Cyclotron 3 Ottignies-Louvain-La-Neuve Belgium
FEI Number
3000256071
Product Code
LHN
Status
Open, Classified
Root Cause
Process design
Initiated
July 10, 2024
Posted
August 21, 2024

Description

IBA Proton Therapy System - PROTEUS 235- Designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation

Reason

Temporary deactivation of collision detection for Forte Robotic Patient Positioning System occasionally experiences issues

Action

IBA isued Urgent Medical Device Correction letter on July 10, 2024. Letter states reason for recall, health risk and action to take: IBA would like to remind you as IBA Proton Therapy System -PROTEUS 235 user of the warnings present in your Clinical User's Guide which emphasize the importance of carefully monitoring the patients behavior during the different treatment phases. IBA would like to also remind you that the manual hand crank could be used by the system operator for the emergency release of the Patient Positioning System in the situations listed in the Clinical User's Guide and Accelerator Operator's Guide and by following the instructions described in both guides. To be signed and given back to IBA representative on site. PTS-8, PTS-10 and PTS-11 versions: Software upgrade IBA will update the software of the Positioning Management System in order that, after a collision, the request to temporarily disable the collision detection will be sent until the status of the collision detection is well disabled. This solution will be deployed at your site by March 2026. PTS-12 versions: users training and labeling change IBA will provide customer training on the system behavior in PTS-12 versions following collision detections. Additionally, IBA will update the labeling to clarify that users can also move the snout or gantry to go out collision situations. These solutions will be deployed at your site by January 2025

Distribution

IL,LA, NJ, MI,OK, TN, TX, VA, WA Foreign: Czech Republic France Germany Italy Japan Poland Russia Sweden

Quantity

19 units