FDA Recall
Terminated
Justright Surgical, JustRight 5mm Reload, REFJR-REL25-2.0, Sterile, EO, Rx Only It is a replacement cartridge used when more than one firing is required during a surgery.
Recall: Z-2691-2014
·
Initiated August 27, 2014
Recall
- Recall Number
- Z-2691-2014
- Event Number
- 69140
- Firm
- Justright Surgical, LLC
- FEI Number
- 3010377594
- Product Code
- GDW
- Status
- Terminated
- Root Cause
- Process design
- Initiated
- August 27, 2014
- Posted
- September 23, 2014
- Terminated
- October 6, 2014
- Address
- 6325 Gunpark Dr, Suite G, Boulder, CO, 80301-3592
Description
Justright Surgical, JustRight 5mm Reload, REFJR-REL25-2.0, Sterile, EO, Rx Only It is a replacement cartridge used when more than one firing is required during a surgery.
Reason
JustRight Surgical is recalling specific lots of JustRight 5mm Stapler and Reload products because they may deliver poor distal staple formation, past the end of the knife cut.
Action
Consignees were notified via letter on August 27, 2014 in addition, the sales rep will hand deliver the Recall Notification on 8/28/14. Acknowledgement Form will be provided and returned products will be sent to STD Med where it will be held in quarantine.
Distribution
Distributed in the states of Colorado, California and New York.
Quantity
3 cases containing 36 units