FUJIFILM Digital Mammography System Aspire HD (FDR MS-1000) with Biopsy Positioner (FDR-1000BPY) The Fujifilm Digital Mammography System, Aspire HD (FDR MS-1000), generates full-field digital mammography images that can, as other full-field digital mammography systems, be used for screening and diagnosis of breast cancer and is intended for use in the same clinical applications as traditional screen-film mammography systems.
Recall
- Recall Number
- Z-2690-2016
- Event Number
- 74525
- Firm
- Fujifilm Medical Systems U.S.A., Inc.
- FEI Number
- 1000513161
- Product Code
- MUE
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- June 9, 2016
- Terminated
- September 19, 2016
- Address
- 419 West Ave, Stamford, CT, 06902-6343
Description
FUJIFILM Digital Mammography System Aspire HD (FDR MS-1000) with Biopsy Positioner (FDR-1000BPY) The Fujifilm Digital Mammography System, Aspire HD (FDR MS-1000), generates full-field digital mammography images that can, as other full-field digital mammography systems, be used for screening and diagnosis of breast cancer and is intended for use in the same clinical applications as traditional screen-film mammography systems.
If an exposure is aborted during a stereo biopsy workflow, an error may occur which would prevent the system from continuing to operate properly.
The firm, Fujifilm, sent a "FIELD SAFETY NOTICE" letter dated June 9, 2016 via Federal Express to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to read and follow the instructions in the letter; call your local service representative, in case an error occurs, and complete and return the Customer Feedback Form via fax (203)-251-7862. Both affected systems will be corrected via on-site visits by FMSU Customer Service Engineers. If you have any questions about this matter, please contact the Product Manager, Women's Health directly at (303)-888-4549 or call your local FUJIFILM office.
US Distribution to: NC and HI.
2 Units