Coulter PrepPlus, Part Number 286600, Product Usage: The COULTER PrepPlus is a microprocessor-controlled pipetting and diluting system designed for automating sample preparation or assay methods. It is capable of aspirating and dispensing liquid samples.
Recall
- Recall Number
- Z-2689-2017
- Event Number
- 77362
- Firm
- Beckman Coulter Inc.
- FEI Number
- 2050012
- Product Code
- PER
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- May 22, 2017
- Terminated
- March 31, 2021
- Address
- 250 S Kraemer Blvd, Brea, CA, 92821-6232
Description
Coulter PrepPlus, Part Number 286600, Product Usage: The COULTER PrepPlus is a microprocessor-controlled pipetting and diluting system designed for automating sample preparation or assay methods. It is capable of aspirating and dispensing liquid samples.
When samples are prepared using the PrepPlus or PrepPlus 2 sample preparation systems and then run on a Flow cytometer, there is an imprecision in the absolute counts on only the first tube in a run.
The recalling firm initiated their recall by letter disseminated via email on 05/22/2017. The letter requested the following actions be taken: 1. Insert an empty daughter tube in place of a patient sample as the first tube in the carousel that is programmed for the addition of Flow-Count Fluorospheres. 2. Refer to the COULTER PrepPlus Operators Guide / PrepPlus 2 Instructions for Use (IFU) for instructions on adding Flow-Count Fluorospheres to daughter tubes. 3. Review all results following the flow cytometer data review instructions referenced in the associated IFU. Beckman Coulter also requested the consignee share this information with their laboratory staff and retain this notification as part of their laboratory Quality System documentation. If the consignee forwarded any of the affected product(s) listed above to another laboratory, They were asked to provide them a copy of the letter.
Worldwide Distribution
43 units