MEVION S250 Proton Radiation Treatment System; intended to deliver proton radiation treatment to patients

Recall: Z-2686-2020 · Initiated June 23, 2020

Recall

Recall Number
Z-2686-2020
Event Number
86028
Firm
Mevion Medical Systems, Inc.
FEI Number
3007087027
Product Code
LHO
Status
Terminated
Root Cause
Device Design
Initiated
June 23, 2020
Terminated
December 9, 2020
Address
300 Foster St, Littleton, MA, 01460-2017

Description

MEVION S250 Proton Radiation Treatment System; intended to deliver proton radiation treatment to patients

Reason

The perforated screen that divides the inside of the process water tank has a very sharp edge across the top of the Heat Exchanger leading to potential injury for servicer.

Action

This screen would not be accessed by normal user (only servicer). Firm is notifying servicing personnel and not consumer/user level. No customer letters/communication sent. Servicer will install edge guard on effected units. Technical Service Bulletin issued to servicer.

Distribution

MO, NJ, OK, FL, OH, DC, Netherlands

Quantity

8 units