FDA Recall Terminated

Terumo Cardiovascular Systems, CDI 100 Hematocrit/Oxygen saturation monitoring system, catalog 100.

Recall: Z-2684-2011 · Initiated April 21, 2011

Recall

Recall Number
Z-2684-2011
Event Number
58532
Firm
Terumo Cardiovascular Systems Corporation
FEI Number
1828100
Product Code
DRY
Status
Terminated
Root Cause
Device Design
Initiated
April 21, 2011
Posted
June 28, 2011
Terminated
August 6, 2012
Address
6200 Jackson Road, Ann Arbor, MI, 48103-9586

Description

Terumo Cardiovascular Systems, CDI 100 Hematocrit/Oxygen saturation monitoring system, catalog 100.

Reason

Inaccurate readings may result after Methylene Blue, an intravascular dye, is administered to the patient. The presence of Methylene Blue may interfere with the readings from the shunt sensor for the CDI 500 system in a way that would cause inaccurate pH values. Because the pH value is used to calculate other values, these parameters could also be affected by the presence of Methylene Blue: K

Action

Terumo sent URGENT MEDICAL DEVICE CORRECTION letters to Chief of Perfusion, Department of Cardiothoracic Surgery or Director of Operating Room Services, dated 4/21/2011. The letter explained the reason for the correction and the potential hazard. The letter stated that Terumo was updating the Operators Manual for all CDI systems and would provide the manuals to all users when available. Consignees with questions should call 1-800-521-2818. Monday- Friday 8 am- 6 pm.

Distribution

Worldwide distribution, including USA, Australia, Barbados, BELGIUM, BRAZIL, CANADA, Chile, Colombia, Costa Rica, Dominican Republic, ECUADOR, El Salvador, France, Germany, Guatemala, Holland, Honduras, Hong Kong, India, Indonesia, Japan, Jordan, Korea, Libya, MALAYSIA, MEXICO, PAKISTAN, PANAMA, Paraguay, Saudi Arabia, SINGAPORE, South Africa, Spain, TAIWAN, THAILAND, The Phillipines, TRINIDAD, UNITED ARAB EMIRATES, Uruguay, Venezuela and Vietnam.

Quantity

1193