FDA Recall Open, Classified

Alinity hq Analyzer, REF: 09P68-01

Recall: Z-2681-2024 · Initiated June 14, 2024

Recall

Recall Number
Z-2681-2024
Event Number
94958
Firm
Abbott Laboratories
FEI Number
2919069
Product Code
GKZ
Status
Open, Classified
Root Cause
Software design
Initiated
June 14, 2024
Posted
August 16, 2024
Address
4551 Great America Pkwy, Santa Clara, CA, 95054-1208

Description

Alinity hq Analyzer, REF: 09P68-01

Reason

1)If Complete Blood Count with differential and reticulocyte(CBC+Diff+Retic) sample is run, then immediately following on same rack CBC+Diff sample is run, then falsely low red blood cell count may occur, generating falsely high Mean Cell Hemoglobin(MCH)/MCHC and falsely low hematocrit results. 2)Cell events may be incorrectly counted as basophil(BASO), resulting in increased BASO and %BASO counts

Action

On 6/14/24, correction notices were sent to consignees informing them of the following: 1) Issue 1: Continue to follow the Alinity h-series Operations Manual to verify flagged results. In addition, you must select one of the two options below depending on your laboratory workflow: Option 1: Batch your CBC+DIFF+RETIC samples on an analyzer. Ensure a Background cycle is run prior to processing the next CBC+DIFF sample on that analyzer. OR Option 2: The MCHC for the laboratories are typically set between 31 36 g/dL. In the event your patient sample flags MCHC results greater than the reference range in your lab, rerun the sample and ensure the rerun does not immediately follow a CBC+DIFF+RETIC sample. 2) Issue 2: The %B high limit is typically set by the Laboratory between 2 5%. In the event your patient sample %B results are greater than the laboratory established high limit, follow the guidelines of your laboratory to rerun the sample or perform a manual slide review. Continue to follow the Alinity h-series Operations Manual to verify flagged results. 3) Provide a copy of the accompanying Product Correction letter to the laboratory manager, supervisor or health professional responsible for the impacted product. 4) If you have forwarded the product listed above to other laboratories, please inform them of this Product Correction and provide to them a copy of this letter. 5) Retain this letter for your laboratory records. 6) Complete and return the Customer Reply Form via email to [email protected] Your Abbott representative will be scheduling a mandatory upgrade of your Alinity hq Analyzer to install software versions(Country): 5.10 (US), 5.9 (Canada and South Korea), 5.8 (OUS except Canada and South Korea). If you or any of the health care providers you serve have questions regarding this information, please contact Customer Service at 1-877-4ABBOTT (available 24 hours a day, 7 days a week).

Distribution

Worldwide - US Nationwide distribution in the states of NY, OK, NC and the countries of ALBANIA, ARGENTINA, AUSTRALIA, BANGLADESH, BELGIUM, BOSNIA AND HERZEGOVINA, Botswana, BRAZIL, BULGARIA, CANADA, CAYMAN ISLANDS, CHILE, COLOMBIA, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, FRANCE, GERMANY, GREECE, HONDURAS, HONG KONG, HUNGARY, INDIA, INDONESIA, ITALY, JAMAICA, JAPAN, JORDAN, KUWAIT, LEBANON, LITHUANIA, MALAYSIA, MALDIVES, MEXICO, NETHERLANDS, NORWAY, OMAN, PAKISTAN, PANAMA, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, SAUDI ARABIA, SERBIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWITZERLAND, THAILAND, TRINIDAD & TOBAGO, TURKEY, UAE, UK, URUGUAY, VIETNAM.

Quantity

816