FDA Recall Terminated

No packaging. The device is a computed tomography scanner. Intended to provide corss sectional images of the human body and visualization of the internal organs of the body.

Recall: Z-2678-2014 · Initiated January 9, 2007

Recall

Recall Number
Z-2678-2014
Event Number
69102
Firm
Philips Medical Systems (Cleveland) Inc
FEI Number
1525965
Product Code
JAX
Status
Terminated
Root Cause
Labeling design
Initiated
January 9, 2007
Posted
September 22, 2014
Terminated
November 6, 2015
Address
595 Miner Rd, Cleveland, OH, 44143-2131

Description

No packaging. The device is a computed tomography scanner. Intended to provide corss sectional images of the human body and visualization of the internal organs of the body.

Reason

When the Extended Display Firled ov View reconstruction is used, the CT numbers and anatomy in teh extended region are not accurate.

Action

Philips sent a Product Safety Notification letter dated December 21, 2006 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to distribute the Product Safety Notification to all affected users and keep a copy of the notification with the Instructions for Use until further notice. Customers with questions were instructed to contact Customer Care Center at 1-800-722-9377, option 5 - Diagnostic Imaging, option 1 - CT. For questions regarding this recall call 440-483-7600.

Distribution

Worldwide Distribution - US including MO, TX, MD, AR, VA, KY, RI, NJ, MA, NY, LA, PA, VT, SC, CA, OH, FL, MN, MI and Internationally to Netherlands, Canada, Finland, Austria, France, Malaysia, Denmark, Norway, UK, Australia, Belgium, New Zealand.

Quantity

61