FDA Recall Terminated

Remel Spectra MRSA, Ref R01821, 10 plates per package, and Ref R01822, 100 plates per package. The firm name on the label is Remel, Lenexa, KS. The product is a selective and differential chromogenic medium recommended for use in the qualitative detection of nasal colonization of methicillin-resistant Staphylococcus aureus (MRSA) to aid in the prevention and control of MRSA in healthcare settings. It is also intended for use in the qualitative detection of MRSA from positive blood cultures demonstrating gram-positive cocci on Gram stain.

Recall: Z-2677-2011 · Initiated May 11, 2011

Recall

Recall Number
Z-2677-2011
Event Number
59005
Firm
Remel Inc
FEI Number
1924669
Product Code
JSO
Status
Terminated
Root Cause
Process control
Initiated
May 11, 2011
Posted
June 28, 2011
Terminated
June 18, 2013
Address
12076 Santa Fe Trail Dr, Lenexa, KS, 66215-3519

Description

Remel Spectra MRSA, Ref R01821, 10 plates per package, and Ref R01822, 100 plates per package. The firm name on the label is Remel, Lenexa, KS. The product is a selective and differential chromogenic medium recommended for use in the qualitative detection of nasal colonization of methicillin-resistant Staphylococcus aureus (MRSA) to aid in the prevention and control of MRSA in healthcare settings. It is also intended for use in the qualitative detection of MRSA from positive blood cultures demonstrating gram-positive cocci on Gram stain.

Reason

The product failed to prevent growth of methicillin-sensitive Staphylococcus aureus (MSSA)

Action

Remel, Inc. sent an "IMPORTANT MEDICAL DEVICE PRODUCT RECALL NOTICE" letter, dated May 10, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter instructs customers to review their inventory and discard all remaining product lots. Additionally, customers are to complete the Product Inventory Checklist and indicate whether or not they desire replacement of the product or credit. A self-addressed , stamped envelope was enclosed in the letter to use to return the checklist. Questions can be answered by contacting the Technical Services Department at 800-255-6730 or 913-888-0939.

Distribution

Nationwide Distribution--including Puerto Rico.

Quantity

187/10-packs and 261/100-packs