Spectra Optia Apheresis System, REF: 61000, CaridianBCT, Lakewood, CO 80215. The Spectra Optia Apheresis System, a blood component separator, is intended for use in therapeutic apheresis applications. The system may be used to perform Therapeutic Plasma Exchange, Mononuclear Cell Collection, and Red Blood Cell Exchange.
Recall
- Recall Number
- Z-2674-2011
- Event Number
- 58909
- Firm
- Caridian BCT, Incorporated
- FEI Number
- 1000117351
- Product Code
- LKN
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- December 9, 2010
- Posted
- June 28, 2011
- Terminated
- September 12, 2011
- Address
- 10811 W Collins Ave, Lakewood, CO, 80215-4440
Description
Spectra Optia Apheresis System, REF: 61000, CaridianBCT, Lakewood, CO 80215. The Spectra Optia Apheresis System, a blood component separator, is intended for use in therapeutic apheresis applications. The system may be used to perform Therapeutic Plasma Exchange, Mononuclear Cell Collection, and Red Blood Cell Exchange.
An additional air detection system is being added to all existing systems to detect air in the system's return line.
CaridianBCT sent a "MEDICAL DEVICE RECALL" letter dated December 2010 to all customers. The letter describes the product, problem, and actions to be taken by the customers. Customers were advised to continue using their systems as long as they followed the instructions in the Operator's Manual. An Acknowledgement of Receipt form was included with the letter to be completed and faxed to 1-303-876-9277 or e-mailed to [email protected]. A CaridianBCT representative will schedule a visit to facilities to upgrade their systems within the next 6 months. Further information is available by contacting the CaridianBCT Support Center at 877-339-4228 or 303-231-4357.
Worldwide Distribution--USA (nationwide) and countries of Australia, Belgium, Canada, China, Hong Kong, Japan, Malaysia, Mexico, Pakistan, Saudi Arabia, Singapore, South Korea, Taiwan, Thailand, and Vietnam.
594 units