FDA Recall Open, Classified

Ashvins variant HEYEX 2 / HEYEX PACS. Software Version: Version 2.6.0 (Build 2088) to 2.6.8 (Build 2220). Device Type: Software (Medical Image Management, Picture Archiving and Communication System)

Recall: Z-2673-2025 · Initiated September 9, 2025

Recall

Recall Number
Z-2673-2025
Event Number
97509
Firm
MedicalCommunications GmbH Max-Jarecki-Str. 8 Heidelberg Germany
FEI Number
3003995303
Product Code
NFJ
Status
Open, Classified
Root Cause
Software design
Initiated
September 9, 2025
Posted
September 30, 2025

Description

Ashvins variant HEYEX 2 / HEYEX PACS. Software Version: Version 2.6.0 (Build 2088) to 2.6.8 (Build 2220). Device Type: Software (Medical Image Management, Picture Archiving and Communication System)

Reason

Potential that the measured value may be smaller than the actual area.

Action

Ashvins notified consignees on about 09/09/2025 via emailed letter. Consignees were instructed to cease using the freeform measurement function in the MMV until the software update has been performed, which is now available. Consignees were instructed to perform the update as soon as possible, complete the digital Field Safety Notice Acknowledgement and Receipt Form, and maintain a copy of the letter and the user manual with the device until the software update is complete.

Distribution

US Nationwide distribution.

Quantity

474 units