SafeStep Huber Needle Set with Y Injection Site and Medegen Needleless Injection Cap; Catalog Numbers LH-0029YN (UDI 00801741066177), LH-0030YN (UDI 00801741066191), LH-0031YN (UDI 00801741066214), LH-0032YN (UDI 00801741066238), LH-0033YN (UDI 00801741066252), LH-0034YN (UDI 00801741066276), LH-0035YN (UDI 00801741066290 ), LH-0036YN (UDI 00801741066313), LH-0037YN (UDI 00801741066337), LH-0038YN (UDI 00801741066351), LH-0039YN (UDI 00801741066313), LH-0040YN (UDI 00801741066399)
Recall
- Recall Number
- Z-2669-2020
- Event Number
- 85853
- Firm
- Becton Dickinson & Company
- FEI Number
- 2243072
- Product Code
- PTI
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- June 19, 2020
- Posted
- July 20, 2020
- Terminated
- July 29, 2022
- Address
- 1 Becton Dr, Franklin Lakes, NJ, 07417-1815
Description
SafeStep Huber Needle Set with Y Injection Site and Medegen Needleless Injection Cap; Catalog Numbers LH-0029YN (UDI 00801741066177), LH-0030YN (UDI 00801741066191), LH-0031YN (UDI 00801741066214), LH-0032YN (UDI 00801741066238), LH-0033YN (UDI 00801741066252), LH-0034YN (UDI 00801741066276), LH-0035YN (UDI 00801741066290 ), LH-0036YN (UDI 00801741066313), LH-0037YN (UDI 00801741066337), LH-0038YN (UDI 00801741066351), LH-0039YN (UDI 00801741066313), LH-0040YN (UDI 00801741066399)
Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.
URGENT MEDICAL DEVICE RECALL notification letters dated 6/19/20 were sent to customers. PLEASE TAKE THE FOLLOWING ACTIONS: 1. Immediately review your inventory for the specific catalog and lot numbers referenced in Attachment A. Discontinue use of those devices and destroy all product subject to this recall following your institutions process for destruction. MDS-20-1979 Page 2 of 28 BD Medical 1 Becton Drive Franklin Lakes, NJ 07417 USA bd.com 2. Share this notification with all users of the product within your facility network to ensure they are also aware of this recall. 3. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification per FDA requirements and subsequently process your credit accordingly for the discarded material. a. Customers who purchased product directly from BD will receive credit. b. Customers who did not purchase product directly from BD will need to contact their distributor to obtain credit. 4. Report any adverse health consequences experienced with the use of this product to BD at [email protected]. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch Phone: 1800FDA1088 (18003321088) Mail: MedWatch, HF2, FDA, 5600 Fishers Lane, Rockville, MD 208529787 Customer/Technical Support Phone: 8002901689 Monday Friday between 8:00 am to 5:00 pm MDT [email protected] An Urgent Medical Device Recall notification letter dated 8/25/20 was sent to customers informing them of two additional lots of affected product. An URGENT MEDICAL DEVICE RECALL UPDATE dated 10/7/20 was sent to customers informing them of two additional lots of affected product.
Worldwide Distribution
5,296,132 total