FDA Recall Open, Classified

Veran : Endobronchial Always on Track Models: INS-0392 AOTT 21ga Needle, 12mm L, 1.8mm OD, 5/BX INS-0382 AOTT 19ga Needle, 12mm L, 1.8 OD, 5/Bx INS-0372 AOTT Serrated Forceps, 1.8mm OD, 5/Bx INS-0362 AOTT Oval Cup Forceps, 1.8mm OD, 5/Bx INS-0352 "Always-On Tip Tracked (AOTT) Brush, 15mm L, 1.8mm OD, 5/Bx" INS-0322 View Peripheral Catheter, 3.2mm OD, 2.0mm WC INS-0305 Always-On Tip Tracked Guidewire, 1.0mm, 950mm INS-0304 Always-On Tip Tracked Guidewire, 1.0mm, 735mm INS-0050 vPad Patient Tracker, 5/Bx INS-0048 vPad Connector Cable

Recall: Z-2664-2023 · Initiated August 10, 2023

Recall

Recall Number
Z-2664-2023
Event Number
92893
Firm
Olympus Corporation of the Americas
FEI Number
2429304
Product Code
JAK
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
August 10, 2023
Posted
September 29, 2023
Address
3500 Corporate Pkwy, Center Valley, PA, 18034-8229

Description

Veran : Endobronchial Always on Track Models: INS-0392 AOTT 21ga Needle, 12mm L, 1.8mm OD, 5/BX INS-0382 AOTT 19ga Needle, 12mm L, 1.8 OD, 5/Bx INS-0372 AOTT Serrated Forceps, 1.8mm OD, 5/Bx INS-0362 AOTT Oval Cup Forceps, 1.8mm OD, 5/Bx INS-0352 "Always-On Tip Tracked (AOTT) Brush, 15mm L, 1.8mm OD, 5/Bx" INS-0322 View Peripheral Catheter, 3.2mm OD, 2.0mm WC INS-0305 Always-On Tip Tracked Guidewire, 1.0mm, 950mm INS-0304 Always-On Tip Tracked Guidewire, 1.0mm, 735mm INS-0050 vPad Patient Tracker, 5/Bx INS-0048 vPad Connector Cable

Reason

Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments) manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and to not interfere with other devices within its operating environment

Action

Olympus Corporation of the Americas ( Olympus ),[Veran Medical Technologies, a wholly owned subsidiary] issued Urgent Medical Device Recall Letter on 8/10/23 to: Endoscopy Department, Pulmonary Department, Risk Management for the VERAN ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments. Letter states reason for recall, health risk and action to take: Veran requests you to take the following actions: 1. Inspect your inventory and identify any Veran models. Please check all areas of the hospital to determine if any of these devices remain in inventory. 2. Call your Olympus customer service representative at 1-800-848-9024, option 2. Olympus will issue a Return Material Authorization to return any affected product at no charge. Your Olympus Customer Service representative will advise you on reimbursement for your Veran devices. 3. Olympus requests that you acknowledge receipt of this letter through the Olympus recall portal. a. Go to https://olympusamerica.com/recall b. Enter the file (recall) number: 0430 4. If you have further distributed this product, identify your customers, forward them this notification, and appropriately document your notification process. If you require additional information, please contact Tara Safi, Field Corrective Actions Administrator at [email protected].

Distribution

Worldwide distribution - US Nationwide and the countries of CN, DE, HK, IT, KR, SG, TH, TW.

Quantity

60102 units