FDA Recall Terminated

EXACTECH Equinoxe Reverse Starter Reamer - Catalog #: 321-05-01

Recall: Z-2663-2017 · Initiated May 25, 2017

Recall

Recall Number
Z-2663-2017
Event Number
77004
Firm
Exactech, Inc.
FEI Number
1038671
Product Code
HTX
Status
Terminated
Root Cause
Error in labeling
Initiated
May 25, 2017
Terminated
July 26, 2019
Address
2320 NW 66th Ct, Gainesville, FL, 32653-1630

Description

EXACTECH Equinoxe Reverse Starter Reamer - Catalog #: 321-05-01

Reason

Instructions for use (IFU) is updated to clarify the proper technique and instrumentation to ream the Glenoid, i.e., to include a caution statement that the surgeon should ream on-axis and that reaming off-axis may result in instrument breakage.

Action

On May 22, 2017 Exactech sent a Notification to all their Direct Accounts an Distributors telling them that a new updated technical techniques LIT # 718-01-30 Rev J is now available in their website (www.exac.com) or can be requested by contacting the Customer Service Representative at 800-392-2832, fax 352-378-2617 or via e-mail at [email protected]. This updated the technical techniques for implantation for Equinoxe Platform Shoulder System to clarify the proper technique for use of the Pilot Tip Reamers, i.e., to include caution statement that the surgeon should avoid applying a bending force to the pilot tip reamer and should avoid using the reamer to retract the humeral head, as this may result in instrument breakage. The notification also request to forward and informed this notice to all the accounts that have received this product and to complete and return the attached Field Advisory Notice Respond Form within 5 days of receipt. For more information please contact: Kaya Davis, Inventory Representative at [email protected] or by phone at (352) 377-1140

Distribution

AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, MA, MD, ME, MN, NC, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WI, Hawaii and Puerto Rico Australia, Austria, Brazil, Canada, China, France, Germany, Greece, India, Italy, Korea, Luxembourg, Singapore, South Africa, Spain, Sweden, Switzerland, The Netherlands and United Kingdom

Quantity

382