Elekta Impac Software Sequencer Verify & Record System, for MOSAIQ The intended use of Sequencer (generically referred to as a verification system) is to assist in the process of process of patient treatment on a radiotherapy treatment machine. It gives the capability to notify users of actions that need to take place prior to treatment, to display reference images for set up purposes and to auto setup the machine to predefined settings.
Recall
- Recall Number
- Z-2662-2011
- Event Number
- 58186
- Firm
- Impac Medical Systems Inc
- FEI Number
- 1000123805
- Product Code
- IYE
- Status
- Terminated
- Root Cause
- Software design (manufacturing process)
- Initiated
- March 10, 2011
- Posted
- June 24, 2011
- Terminated
- September 16, 2011
- Address
- 100 Mathilda Pl, Fl 5th, Sunnyvale, CA, 94086-6076
Description
Elekta Impac Software Sequencer Verify & Record System, for MOSAIQ The intended use of Sequencer (generically referred to as a verification system) is to assist in the process of process of patient treatment on a radiotherapy treatment machine. It gives the capability to notify users of actions that need to take place prior to treatment, to display reference images for set up purposes and to auto setup the machine to predefined settings.
There's a problem in the database conversion program that is used when upgrading from the Multi-ACCS to MOSAIQ or from one version of MOSAIQ to another. If the gantry start angle is incorrect prior to pressing beam-on, the patient will be mistreated.
The firm, Elekta, sent a "USER NOTICE" to its customers with an attached confirmation of receipt form. The notice describes the product, problem and actions to be taken. The customers were instructed to follow the Interim mitigation instructions noted in the notice -verify the Arc start angle of the Simple ARC with MLC treatment field definition prior to delivery and if the customer identifes a field with an errant gantry start angle, to cease using that plan immediately and consult with the original plan for that field and make the correction to the gantry start angle. In addition, the customers were instructed to complete and return the Confirmation of Receipt via fax to: North America and Asia at 702-992-5002 or Europe and Africa +44 (0)1293 654401, Attn: Elekta OIS Su[pport and/or scan and email to: [email protected], [email protected] or by mail; and distribute this notice to any all users of IMPAC software at their organization who are potentially affected by this issue. For questions or requests for upgrade, email [email protected] (North America/Asia) or [email protected] (Europe) or call USA 1-800-488-4672.
Worldwide distribution: USA (nationwide) and country of Canada.