FDA Recall Terminated

Elekta Impac Software Sequencer Verify & Record System, for MOSAIQ The intended use of Sequencer (generically referred to as a verification system) is to assist in the process of process of patient treatment on a radiotherapy treatment machine. It gives the capability to notify users of actions that need to take place prior to treatment, to display reference images for set up purposes and to auto setup the machine to predefined settings.

Recall: Z-2662-2011 · Initiated March 10, 2011

Recall

Recall Number
Z-2662-2011
Event Number
58186
Firm
Impac Medical Systems Inc
FEI Number
1000123805
Product Code
IYE
Status
Terminated
Root Cause
Software design (manufacturing process)
Initiated
March 10, 2011
Posted
June 24, 2011
Terminated
September 16, 2011
Address
100 Mathilda Pl, Fl 5th, Sunnyvale, CA, 94086-6076

Description

Elekta Impac Software Sequencer Verify & Record System, for MOSAIQ The intended use of Sequencer (generically referred to as a verification system) is to assist in the process of process of patient treatment on a radiotherapy treatment machine. It gives the capability to notify users of actions that need to take place prior to treatment, to display reference images for set up purposes and to auto setup the machine to predefined settings.

Reason

There's a problem in the database conversion program that is used when upgrading from the Multi-ACCS to MOSAIQ or from one version of MOSAIQ to another. If the gantry start angle is incorrect prior to pressing beam-on, the patient will be mistreated.

Action

The firm, Elekta, sent a "USER NOTICE" to its customers with an attached confirmation of receipt form. The notice describes the product, problem and actions to be taken. The customers were instructed to follow the Interim mitigation instructions noted in the notice -verify the Arc start angle of the Simple ARC with MLC treatment field definition prior to delivery and if the customer identifes a field with an errant gantry start angle, to cease using that plan immediately and consult with the original plan for that field and make the correction to the gantry start angle. In addition, the customers were instructed to complete and return the Confirmation of Receipt via fax to: North America and Asia at 702-992-5002 or Europe and Africa +44 (0)1293 654401, Attn: Elekta OIS Su[pport and/or scan and email to: [email protected], [email protected] or by mail; and distribute this notice to any all users of IMPAC software at their organization who are potentially affected by this issue. For questions or requests for upgrade, email [email protected] (North America/Asia) or [email protected] (Europe) or call USA 1-800-488-4672.

Distribution

Worldwide distribution: USA (nationwide) and country of Canada.

Quantity

840