Nussloch GmbH ASP6025 Tissue Processor
Recall
- Recall Number
- Z-2660-2017
- Event Number
- 77603
- Firm
- Leica Microsystems, Inc.
- FEI Number
- 3000210133
- Product Code
- IEO
- Status
- Open, Classified
- Root Cause
- Labeling mix-ups
- Initiated
- June 19, 2017
- Address
- 1700 Leider Ln, Buffalo Grove, IL, 60089-6622
Description
Nussloch GmbH ASP6025 Tissue Processor
Incorrect labeling for specified voltage for the Alarm Connectors on the rear side of the unit.
Customers were notified via email on 06/19/2017. Instructions to customers included to clean surface around the Alarm Connectors and place the newly provided label at the Alarm Connectors (over the engraved specification). They were also instructed to update their current Instructions For Use and update the online help function in the software, as well as return the response form.
Domestic: AR, AZ, CA, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MN, MO, MT, NC, NE, NJ, NM, NV, NY, OH, PA, SC, TN, TX, UT, VA, WA, WI, WV, & DC. Foreign: Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, England, France, Georgia, Germany, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Korea, Kuwait, Lithuania, Malaysia, Maldives, Mexico, Morocco, Myanmar, Netherlands, Norway, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Switzerland, Taiwan, Thailand, Turkey, & United Arab Emirates.
812 units