FDA Recall Terminated

Precision Ice One-Patient-Use Control Unit and Wrap System (Model # 4BB02BK). The device is intended to reduce edema and promote healing of soft-tissue injuries and lymphedema through stimulation of the lymphatic system. The device is a therapeutic system which consists of a control unit and fluid circulating pads. The pads are mounted in wraps designed for specific parts of the body such as ankle, knee, elbow, and shoulder. The pads provide cold treatment and pressure which help promote healing. The owner's manual is labeled in part: "PRO TRAINERS' CHOICE COMPANY, INC.***www.precisionice.com***[email protected]***Mailing Address PO Box 27***Kingston, WA 98346 USA***Phone: 360-297-3902***Shipping Address***26129 Calvary Lane NE, Suite 200***Kingston, WA 98346 USA***Fax: 360-297-3903". The sticker which is placed on the back of the control board is labeled in part: "MOD#4BB02BK***SER#100***". The shipping box is labeled in part: "Pro Trainers' Choice' LLC***Model #4BB02BK***Serial #100____***". The knee wrap is labeled in part: "***Pro Trainers' Choice Co. Inc./Precision Ice One Patient Use Knee Wrap***#1BK01051KN***This device is to be used for one patient only; do not re-use!***".

Recall: Z-2659-2011 · Initiated April 18, 2011

Recall

Recall Number
Z-2659-2011
Event Number
58770
Firm
Pro Trainers' Choice Company
FEI Number
3004733525
Product Code
ILO
Status
Terminated
Root Cause
Other
Initiated
April 18, 2011
Posted
June 24, 2011
Terminated
March 1, 2012
Address
26129 Calvary Lane NE PO BOX 27, Suite 200, Kingston, WA, 98346

Description

Precision Ice One-Patient-Use Control Unit and Wrap System (Model # 4BB02BK). The device is intended to reduce edema and promote healing of soft-tissue injuries and lymphedema through stimulation of the lymphatic system. The device is a therapeutic system which consists of a control unit and fluid circulating pads. The pads are mounted in wraps designed for specific parts of the body such as ankle, knee, elbow, and shoulder. The pads provide cold treatment and pressure which help promote healing. The owner's manual is labeled in part: "PRO TRAINERS' CHOICE COMPANY, INC.***www.precisionice.com***[email protected]***Mailing Address PO Box 27***Kingston, WA 98346 USA***Phone: 360-297-3902***Shipping Address***26129 Calvary Lane NE, Suite 200***Kingston, WA 98346 USA***Fax: 360-297-3903". The sticker which is placed on the back of the control board is labeled in part: "MOD#4BB02BK***SER#100***". The shipping box is labeled in part: "Pro Trainers' Choice' LLC***Model #4BB02BK***Serial #100____***". The knee wrap is labeled in part: "***Pro Trainers' Choice Co. Inc./Precision Ice One Patient Use Knee Wrap***#1BK01051KN***This device is to be used for one patient only; do not re-use!***".

Reason

The control panel of the Precision Ice One Patient Use Control Unit and Wrap System has a overheating issue. The control panel printed circuit boards were missing a jumper connection across resistors R28 A, B, C, and D.

Action

Pro Trainers' Choice sent a Mandate-Medical Device Recall on April 18, 2011 and April 20, 2011 to all affected costumers. The letter included a description of the product, affected model and lots. It instructed the costumers to discontinue using the Precision Ice, Model 4BBK02BK until the device is returned and corrected by Pro Trainers' Choice. The device should be returned to PO Box 27, Kingston, WA 98346. The customers were instructed to notify their sub-account customers of this recall if the devices were further distributed. Customers can call Pro Trainers' Choice Company at 360-297-3902 or visit the website www.precisionice.com for any questions about this recall.

Distribution

Nationwide distribution including CA, CO, FL, IN, KS, NY, OR, TX, and WI.

Quantity

336 units