FDA Recall Terminated

Laparoscopic Ultrasound Transducer (LAP Transducer) is used on the SonoSite MicroMaxx ultrasound system. The LAP Transducer label is labeled in part: "MicroMaxx LAP...Manufactured for SonoSite, Inc., Bothell, WA 98021".

Recall: Z-2654-2010 · Initiated November 2, 2007

Recall

Recall Number
Z-2654-2010
Event Number
49387
Firm
Sonosite, Inc.
FEI Number
3002087856
Product Code
IYN
Status
Terminated
Root Cause
Process control
Initiated
November 2, 2007
Posted
October 1, 2010
Terminated
October 5, 2010
Address
21919 30th Dr Se, Bothell, WA, 98021-3904

Description

Laparoscopic Ultrasound Transducer (LAP Transducer) is used on the SonoSite MicroMaxx ultrasound system. The LAP Transducer label is labeled in part: "MicroMaxx LAP...Manufactured for SonoSite, Inc., Bothell, WA 98021".

Reason

Sterilization of the SonoSite LAP surgical Transducer with the Sterrad NX System may result in damage to the transducer.

Action

On November 2, 2007, SonoSite sent out the Important Information Regarding Sterrad NX Incompatibility letter and the Technical Bulletin 005/07 Rev A to their consignees. The letter advises the consignees that SonoSite has recently found that sterilization of the SonoSite SLT and LAP surgical ultrasound transducers with the Sterrad NX system may result in damage to the transducer. The Technical Bulletin 005/07 Rev B dated November 2, 2007 informs the consignees that the damage has been observed primarily on the circuit board in the transducer connector, and cable and laparoscope damage may also be occurring. All damages have occurred after processing using the Sterrad NX systems in its "Advanced" cycle and the potential damage from the use of "normal" cycle is being evaluated by the firm. Consignees should call the Technical Support at 1-425-951-1330 or 1-877-657-8118 or e-mail to [email protected] for any questions regarding the recall.

Distribution

Devices were distributed to medical centers and hospitals in AR, CA, IN, KS, MA, MD, MI, NC, NE, NJ, NV, OH, SC, TN, TX, WI, Washington DC and to FOREIGN COUNTRIES Australia, China, France, Germany, Greece, Hong Kong, India, Iraq, Italy, Korea, Republic of, Libyan Arab Jamahiriya, Mexico, Spain, United Arab Emirates, United Kingdom, and United States.

Quantity

16 units