Brilliance CT Big Bore Oncology, Model number 728243 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.
Recall
- Recall Number
- Z-2653-2016
- Event Number
- 74820
- Firm
- Philips Medical Systems (Cleveland) Inc
- FEI Number
- 1525965
- Product Code
- JAK
- Status
- Terminated
- Root Cause
- Software Design Change
- Initiated
- July 20, 2016
- Terminated
- July 2, 2018
- Address
- 595 Miner Rd, Cleveland, OH, 44143-2131
Description
Brilliance CT Big Bore Oncology, Model number 728243 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.
After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at iso-center (0,0). If the region of interest is not visualized in the images, a CT rescan may be performed.
Customers were sent a letter dated July 28, 2016 informing them of the problem, under what circumstances it can occur and the actions that the customer can take to minimize the effect of the problem. The letter also includes the actions planned by Philips to correct the problem. Philips plans on releasing a field change order (FCO) to correct the affected systems. If you need any further information or support concerning this issue, contact your local Philips representative or local Philips Healthcare office. For North America and Canada, contact the Customer Care Solutions Center (1 -800-722-9377).