Puritan Bennett Cuff Pressure Manager, REF: 180-03
Recall
- Recall Number
- Z-2651-2023
- Event Number
- 92941
- Firm
- Covidien
- FEI Number
- 2936999
- Product Code
- BSK
- Status
- Open, Classified
- Root Cause
- Component design/selection
- Initiated
- August 22, 2023
- Posted
- September 29, 2023
- Address
- 6135 Gunbarrel Ave, Boulder, CO, 80301-3214
Description
Puritan Bennett Cuff Pressure Manager, REF: 180-03
Inadequate internal fixation of the power inlet port to the cuff pressure manager device causes the external power inlet to became dislodged within the device housing after several uses, which prevents the device from being charged, and could lead to treatment delay.
On 8/22/23 Medtronic mailed recall notices to risk managers and directors of respiratory care informing them of the following: 1) Discontinue use, quarantine, and return affected devices to the firm. 2) Complete and return the customer confirmation form via email to it to [email protected]. 3) Forward the notice to all those who need to be aware within your organization or to any organization where the potentially affected product with the specified serial numbers has been transferred or distributed. If you have any questions regarding this communication, contact customer service at 800-962-9888, Option 2. Quality problems can be reported to Quality Assurance at 800-255-6774 option 4.
US: AL, NJ, MA, MI, NY, NC, CO, TX, FL, KY, UT, CA, AZ, KS, IL, TN. OUS: El Salvador, Sweden, Canary Islands
408