DRX-Revolution Mobile X-Ray System; MODEL DRXR-1; Catalog # 1019397 (all other countries) and Catalog 1060177 (China only) -- COMMON/USUAL NAME: DRX-Revolution --- The DRX-Revolution is a mobile system used to generate and control X-Rays for diagnostic procedures
Recall
- Recall Number
- Z-2651-2017
- Event Number
- 77250
- Firm
- Carestream Health Inc
- FEI Number
- 1315356
- Product Code
- IZL
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- May 25, 2017
- Posted
- June 19, 2017
- Terminated
- June 13, 2018
- Address
- 150 Verona St, Rochester, NY, 14608-1733
Description
DRX-Revolution Mobile X-Ray System; MODEL DRXR-1; Catalog # 1019397 (all other countries) and Catalog 1060177 (China only) -- COMMON/USUAL NAME: DRX-Revolution --- The DRX-Revolution is a mobile system used to generate and control X-Rays for diagnostic procedures
Carestream found a single unit where four screws in the assembly were not applied in the manufacturing process. This could result in separation of the tube and yoke assembly.
Carestream Health sent the URGENT MEDICAL DEVICE CORRECTION Letters and Response Forms to the US consignees on May 25, 2017, via Certified Mail, Certified Receipt Requested. Letters to foreign accounts were distributed via e-mail or courier about the same time. Customers with questions were instructed to call 1-800-328-2910.
Worldwide Distribution - US (nationwide) Internationally to Australia, Brazil, Canada, Chile, China, Colombia, Dominican Republic, India, Japan, Israel, Korea, Malaysia, Maldives, Mexico, New Zealand, Peru, Puerto Rico, Singapore, Switzerland, Taipei, Thailand, Turkey, and UAE.
Domestic: 1449 units; Foreign: 1416 units