FDA Recall Terminated

DRX-Revolution Mobile X-Ray System; MODEL DRXR-1; Catalog # 1019397 (all other countries) and Catalog 1060177 (China only) -- COMMON/USUAL NAME: DRX-Revolution --- The DRX-Revolution is a mobile system used to generate and control X-Rays for diagnostic procedures

Recall: Z-2651-2017 · Initiated May 25, 2017

Recall

Recall Number
Z-2651-2017
Event Number
77250
Firm
Carestream Health Inc
FEI Number
1315356
Product Code
IZL
Status
Terminated
Root Cause
Process control
Initiated
May 25, 2017
Posted
June 19, 2017
Terminated
June 13, 2018
Address
150 Verona St, Rochester, NY, 14608-1733

Description

DRX-Revolution Mobile X-Ray System; MODEL DRXR-1; Catalog # 1019397 (all other countries) and Catalog 1060177 (China only) -- COMMON/USUAL NAME: DRX-Revolution --- The DRX-Revolution is a mobile system used to generate and control X-Rays for diagnostic procedures

Reason

Carestream found a single unit where four screws in the assembly were not applied in the manufacturing process. This could result in separation of the tube and yoke assembly.

Action

Carestream Health sent the URGENT MEDICAL DEVICE CORRECTION Letters and Response Forms to the US consignees on May 25, 2017, via Certified Mail, Certified Receipt Requested. Letters to foreign accounts were distributed via e-mail or courier about the same time. Customers with questions were instructed to call 1-800-328-2910.

Distribution

Worldwide Distribution - US (nationwide) Internationally to Australia, Brazil, Canada, Chile, China, Colombia, Dominican Republic, India, Japan, Israel, Korea, Malaysia, Maldives, Mexico, New Zealand, Peru, Puerto Rico, Singapore, Switzerland, Taipei, Thailand, Turkey, and UAE.

Quantity

Domestic: 1449 units; Foreign: 1416 units