FDA Recall Terminated

First Breath PEDIATRIC Medium Concentration Oxygen Mask with Tubing, 213cm (7ft) (Non-latex) 50/CA Reorder Number: 001471 Oxygen/Aerosol Mask For acute use, e.g. Post-op recovery, minimal trauma, patient transfer; used for patients requiring more oxygen than is delivered via a standard nasal cannula (>6L/minute). During inspiration, the patient draws gases both from the oxygen flowing into the mask through the tubing as well as the room via ports on the sides of the mask. Oxygen concentration delivered is 40 60% depending on the patients breathing rate and tidal volume

Recall: Z-2651-2011 · Initiated February 11, 2011

Recall

Recall Number
Z-2651-2011
Event Number
57906
Firm
Smiths Medical Hythe Hythe, Kent United Kingdom
FEI Number
3002807908
Product Code
BYG
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
February 11, 2011
Posted
June 24, 2011
Terminated
July 2, 2018

Description

First Breath PEDIATRIC Medium Concentration Oxygen Mask with Tubing, 213cm (7ft) (Non-latex) 50/CA Reorder Number: 001471 Oxygen/Aerosol Mask For acute use, e.g. Post-op recovery, minimal trauma, patient transfer; used for patients requiring more oxygen than is delivered via a standard nasal cannula (>6L/minute). During inspiration, the patient draws gases both from the oxygen flowing into the mask through the tubing as well as the room via ports on the sides of the mask. Oxygen concentration delivered is 40 60% depending on the patients breathing rate and tidal volume

Reason

A sticky substance on the inside and/ or outside of the Mask

Action

Smiths Medical notified consignees by letter on 2/14/11 titled Urgent Medical Device Correction Notice via U.S. Mail (USA only) This Notice includes an Urgent Medical Device Correction Notice Confirmation Form that provides instructions to consignees to return unused affected product to Smiths Medical and to complete the effectivity tracking form. On 3/17/11 Smiths Medical issued an additional recall letter to extend the recall to 3 additional lots not included in the original recall letter dated 2/14/11. Also, the firm issued a recall letter on 3/17/11 to include the Value Pack (VP1302) General Anesthesia Kit to include 7 kit lots. Contact Recall Coordinator 1-800-348-6064 prompt 4, ext 2273

Distribution

Worldwide Distribution - USA (nationwide) and the countries of United Arab Emirates, Armenia, Brazil, Canada, France, Hungary, Italy, Jordan, Japan, Libya, Poland, Russian Federation, and Saudi Arabia.

Quantity

1952 units