FDA Recall Open, Classified

ZAP-X Radiosurgery System, Model: 300150

Recall: Z-2639-2025 · Initiated July 25, 2025

Recall

Recall Number
Z-2639-2025
Event Number
97413
Firm
Zap Surgical Systems
FEI Number
3012039133
Product Code
IYE
Status
Open, Classified
Root Cause
Software change control
Initiated
July 25, 2025
Posted
September 26, 2025
Address
590 Taylor Way, San Carlos, CA, 94070-6307

Description

ZAP-X Radiosurgery System, Model: 300150

Reason

If the radiosurgery system triggers a proximity error message during a long gantry move (greater-than180 degrees), and a subsequent proximity error message occurs after the automatic reduction of speed, a software defect could potentially lead the collimator to collide with patient shoulders or the patient table.

Action

On 7/25/2025, correction notices were sent to customers who were asked to do the following: Firm will be revising the software to address this issue. Until we supply you with the revised software, the firm strongly advises that you follow these steps to prevent potential collisions. 1. In the rare event that the system triggers a proximity error message during a long gantry move (greater-than-180-degrees), and a subsequent proximity error message occurs after the automatic reduction of speed, please manually extract the patient. 2. After the patient has been safely extracted, please return the gantry to its home position before resuming any treatment procedures. If you have any further questions related to this situation email [email protected]

Distribution

Worldwide - US Nationwide distribution in the states of FL, CO, NJ, NY and the countries of Germany, France, Spain, Poland, Republic of Korea, T¿rkiye, Paraguay.

Quantity

17