3M Attest Super Rapid 5 Steam-Plus Steam Challenge Pack, REF 41482V; designed to qualify and monitor dynamic-air removal steam sterilization processes
Recall
- Recall Number
- Z-2639-2023
- Event Number
- 92915
- Firm
- 3M Company Health Care Business 3m Center
- FEI Number
- 2110898
- Product Code
- FRC
- Status
- Open, Classified
- Root Cause
- Process control
- Initiated
- August 22, 2023
- Posted
- September 26, 2023
- Address
- 2510 Conway Ave, Bldg 275-5w-06, Saint Paul, MN, 55144-0001
Description
3M Attest Super Rapid 5 Steam-Plus Steam Challenge Pack, REF 41482V; designed to qualify and monitor dynamic-air removal steam sterilization processes
Specific lots of 3M Attest Super Rapid Readout Biological Indicators (1492V), 3M Attest Super Rapid 5 Steam-Plus Challenge (41482V, 41482VF), and 3M Attest Super Rapid Readout Steam Challenge (1496V) may contain biological indicators with a cap that melts, deforms and/or cracks following a steam sterilization cycle of 132C and 135C.
3M issued an Urgent product recall notice to its consignees on 08/24/2023 by email and US mail. The notice explained the problem with the product and requested the following: Discontinue distribution of the product and dispose of the product. Distributors were directed to notify their customers. The firm is offering replacement with the return of the completed response form.
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Canada, China, Colombia, Hong Kong, Italy, Japan, Peru, Poland, South Korea, Spain, Taiwan.
172,464 devices