FDA Recall Terminated

ProVue Instrument Driver for JResultNet (ProVue.jrm) v. 1.5.3 and below JResultNet is a stand-alone software package that provides connectivity between in-vitro diagnostic instruments (IVDs) and Laboratory Information Systems (LIS) aiding the flow of laboratory information, including test orders and results, specimen information, and patient information.

Recall: Z-2637-2014 · Initiated August 15, 2014

Recall

Recall Number
Z-2637-2014
Event Number
69086
Firm
Data Innovations, LLC
FEI Number
3007289202
Product Code
JQP
Status
Terminated
Root Cause
Software design
Initiated
August 15, 2014
Posted
September 12, 2014
Terminated
March 4, 2015
Address
2914 Cleveland Ave, Fort Myers, FL, 33901-6003

Description

ProVue Instrument Driver for JResultNet (ProVue.jrm) v. 1.5.3 and below JResultNet is a stand-alone software package that provides connectivity between in-vitro diagnostic instruments (IVDs) and Laboratory Information Systems (LIS) aiding the flow of laboratory information, including test orders and results, specimen information, and patient information.

Reason

Potential issues with the ProVue driver have been identified which may cause data to be misassociated with patient results under a given set of circumstances.

Action

Data Innovations LLC sent an Important Medical Device Information letter on August 15, 2014, to consignees for the JResultNet using ProVue Instrument Driver JResultNet (provue.jrm) v. 1.5.3 and below. Potential issues with the ProVue driver have been identified which may cause data to be mis-associated with patient results under a given set of circumstances. The user on the ProVue transmits the same results twice back to back by clicking the Send to Host button twice. The ProVue sends two ASTM messages back to back before ending transmission (sending an EOT character) The solution was to disconnect JResultNet from the ProVue instrument and apply the patch to driver provue.jrm when it is available. If you have further questions regarding this notice, please feel free to contact us at [email protected] or 239-931-6004 Toll-Free (from the US and Canada): 1-800-332-0499.

Distribution

Nationwide Distribution including AL, AZ, CA, CT, DC, FL, GA, IN, LA, MI, MO, NC, NH, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, and WA.

Quantity

84