FDA Recall Terminated

1500T12 Cool Path Ablation System - Generator The generator is a microprocessor-controlled device that produces a continuous unmodulated radiofrequency (RF) output of 485 kHz. The generator has a maximum power output of 100 Watts. The catheter delivers the RF power from the generator in a monopolar mode between its distal electrode (tip electrode) and a large indifferent electrode (dispersive pad or "DIP"), which is in compliance with ANSUAAMI Standard HF-18-60601-2-2. The Generator operates in Temperature Control mode only. The generator is a temperature controlled system, where temperature measured by the temperature sensor in the Therapy 1300 Series Steerable Ablation Catheter or other compatible catheters is monitored and the power delivered by the generator adjusts within the selected limits until the desired temperature is achieved. The generator's thermocouple temperature control circuitry is based on a Type T thermocouple. The front panel displays the actual real-time power output, impedance, duration, and measured tip electrode temperature. The amount and duration of RF power delivery is user selectable, The desired catheter tip electrode temperature is also user selectable. A low pass filter is enabled during ablation to permit recording of intracardiac electrograms and alternately disables the filter during stimulation (pacing). The output power of the Generator will shut off if the measured temperature exceeds 80C or if the measured temperature exceeds the user selected temperature set point by more than 5C for more than 3 seconds. The Generator has built-in safety features, which include a self test at power up and automatic RF power shut off if the measured tissue impedance falls below 50 Ohms or exceeds 300 Ohms or the preset impedance value for more than 2 seconds. The 1500Tl2 Cardiac Ablation Generator is only compatible with the Cool Point Irrigation Pump.

Recall: Z-2630-2011 · Initiated May 9, 2011

Recall

Recall Number
Z-2630-2011
Event Number
58859
FEI Number
3007562768
Status
Terminated
Root Cause
Other
Initiated
May 9, 2011
Posted
June 23, 2011
Terminated
April 3, 2012
Address
St Jude Medical One St Jude Medical Dr, Saint Paul, MN, 55117

Description

1500T12 Cool Path Ablation System - Generator The generator is a microprocessor-controlled device that produces a continuous unmodulated radiofrequency (RF) output of 485 kHz. The generator has a maximum power output of 100 Watts. The catheter delivers the RF power from the generator in a monopolar mode between its distal electrode (tip electrode) and a large indifferent electrode (dispersive pad or "DIP"), which is in compliance with ANSUAAMI Standard HF-18-60601-2-2. The Generator operates in Temperature Control mode only. The generator is a temperature controlled system, where temperature measured by the temperature sensor in the Therapy 1300 Series Steerable Ablation Catheter or other compatible catheters is monitored and the power delivered by the generator adjusts within the selected limits until the desired temperature is achieved. The generator's thermocouple temperature control circuitry is based on a Type T thermocouple. The front panel displays the actual real-time power output, impedance, duration, and measured tip electrode temperature. The amount and duration of RF power delivery is user selectable, The desired catheter tip electrode temperature is also user selectable. A low pass filter is enabled during ablation to permit recording of intracardiac electrograms and alternately disables the filter during stimulation (pacing). The output power of the Generator will shut off if the measured temperature exceeds 80C or if the measured temperature exceeds the user selected temperature set point by more than 5C for more than 3 seconds. The Generator has built-in safety features, which include a self test at power up and automatic RF power shut off if the measured tissue impedance falls below 50 Ohms or exceeds 300 Ohms or the preset impedance value for more than 2 seconds. The 1500Tl2 Cardiac Ablation Generator is only compatible with the Cool Point Irrigation Pump.

Reason

The recall was initiated because St. Jude Medical has confirmed its T14 and T12 RF Generator may malfunction upon encountering an unexpected electric current (e.g. static electricity). The resulting malfunction results in a failure of the generator to respond to key presses properly, and may cause unexpected energy delivery or inability to discontinue energy delivery with the start / stop butt

Action

The firm, St. Jude Medical (SJM), sent a "Product Correction" letter dated May 9, 2011 to it's affected customers. The letter described the product, problem and actions to be taken. The customers were instructed to discontinue use of the T14/T12 generators and remote controls until the units have been properly serviced; complete and return the attached field action form, and utilize the units in accordance with the Operators Manual. In case of system malfunction, the customers were instructed to shut off the RF power delivery by turning off the generator via the rocker switch, dislodging the power cord, or disconnecting of the patient cable or indifferent electrode connection. An SJM sales representative will contact the customers in the near future to facilitate repair or exchange of the impacted unit. If you have questions regarding this notice, contact the Marketing Manager, AF at (905-286-4015), Marketing Director, AF/CRM, at (905-286-4107); Product Manager (Tokyo) 03-6255-5740, and/or your SJM sales representative.

Distribution

International distribution only: Canada and Japan.

Quantity

154